Initial Setup
N-595
before performing tests 1 through 4. The following is a brief description of each
test:
Test Setup — This procedure establishes the baseline for all the other
tests. The Test Setup procedure must be performed before performing any
or all of the SRC-MAX tests.
Test 1: BPM — This procedure simulates an O
patient indicating 60 BPM and 200 BPM. The test setup procedure sets up
Test 1 for 60 BPM.
Test 2: SpO
— This procedure simulates an O
2
patient, indicating 75 percent blood oxygen saturation and 90 percent
blood oxygen saturation. The test setup procedure sets up Test 2 for 75
percent blood oxygen saturation.
Test 3: Modulation — This procedure simulates an O
attached to a patient indicating low and high pulse strength. The test setup
procedure sets up Test 3 for low pulse strength.
Test 4: Light — This procedure simulates an O
patient indicating low and high light level passing through the patient at
the sensor site. The test setup procedure sets up Test 4 for low light level.
1. With the monitor turned off, connect the DOC-10 pulse oximetry cable to the
sensor port.
2. Connect the SRC-MAX tester to the other end of the DOC-10 cable.
3. Turn on the monitor by pressing the ON/STANDBY button.
Performance Verification
M
XI
M
XI
M
XI
1.Sensor Port
sensor attached to a
AX
sensor attached to a
AX
M
sensor
XI
AX
sensor attached to a
AX
27