Replacement; Figure 9: Power Supply; Table 11: Power Supply Lead Connections - Nellcor OxiMax N-595 Service Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

Replacement

N-595
7. Lift the Power Supply out of the bottom case.
8. Reconnect the leads to the PEM following the instructions in Table 11, and
Figure 8.

Table 11: Power Supply Lead Connections

Wire Color / Label
Green & Yellow
Brown/Labeled "L"
Blue/Labeled "N"
Red/Labeled "+"
Black/Labeled "-"
9. Place the Power Supply in the bottom case.

Figure 9: Power Supply

Connect To
Equipotential Lug
"L" on the Power Entry Module
"N" on the Power Entry Module
Positive Battery Terminal
Negative Battery Terminal
Disassembly Guide
71

Advertisement

Table of Contents
loading