Replacement
N-595
7. Lift the Power Supply out of the bottom case.
8. Reconnect the leads to the PEM following the instructions in Table 11, and
Figure 8.
Table 11: Power Supply Lead Connections
Wire Color / Label
Green & Yellow
Brown/Labeled "L"
Blue/Labeled "N"
Red/Labeled "+"
Black/Labeled "-"
9. Place the Power Supply in the bottom case.
Figure 9: Power Supply
Connect To
Equipotential Lug
"L" on the Power Entry Module
"N" on the Power Entry Module
Positive Battery Terminal
Negative Battery Terminal
Disassembly Guide