Nellcor OxiMax N-595 Service Manual page 27

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

N-595
The plethysmograph waveform tracks your pulse rate.
The pulse tone is heard.
Your SpO
and pulse rate are flashing in the %SpO
2
The audible alarm sounds, indicating that both parameters have violated
the alarm limits.
13. Press and hold the ALARM SILENCE button until the BPM display indicates
"SEC." Continue to press the ALARM SILENCE button and press the
ADJUST DOWN button until "60" is displayed in the %SpO
14. Press the ALARM SILENCE button.
15. With the monitor's alarm silenced, verify the following:
The alarm remains silenced for 60 seconds.
The ALARM SILENCE indicator lights.
The %SpO
and BPM displays continue to flash.
2
The pulse tone is still audible.
The audible alarm returns in approximately 60 seconds.
16. Press and hold the ALARM SILENCE button until the BPM display indicates
"SEC." Continue to press the ALARM SILENCE button and press the
ADJUST DOWN button until "30" is displayed in the %SpO
17. Press the ADJUST UP button and verify that the displays indicate 60 SEC,
90 SEC, 120 SEC, and OFF. Release the ADJUST UP button when the
display indicates "OFF."
18. Press and release the ALARM SILENCE button. Verify that the monitor's
ALARM SILENCE indicator flashes.
19. Wait approximately 3 minutes. Verify that the monitor's alarm does not
return. After 3 minutes, the monitor's alarm silence reminder beeps three
times, and will continue to do so at approximately 3-minute intervals.
Performance Verification
and BPM displays.
2
display.
2
display.
2
19