Other Messages - Nellcor OxiMax N-595 Service Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

Troubleshooting

Other Messages

60
Error
Meaning
Code
523
The operation just attempted was not completed successfully - for example,
Institutional Defaults could not be reset
524
An unexpected value was received - for example, an out-of-range parameter
was passed to a function
525
EEPROM CRC failure
526
SpO
module not responded
2
527
Institutional parameters lost - e.g. for UIF: Institutional EEPROM section
CRC corrupt
528
Current settings lost - e.g. for UIF: Institutional EEPROM section CRC
corrupt
529
Critical low battery
530
Low battery error
531
External watchdog failure
532
Power PC watchdog failure
533
Boot NVROM uninitialized error
534
Failed CRC check of application code in flash
535
Failed periodic ram CRC check on application code running in RAM
562
SpO
front end reset
2
563
SpO
reported error
2
564
Clinical mode was exited after input was received
565
Communication failures between software modules
566
Excessive resets before UIF runs
567
An unexpected interrupt has been asserted
568
General failure in UIF module generic post
569
BOOT application program is corrupt - CRC does not match
570
RTC was restarted
574
Excessive restarts within 1 minute
In addition to the error codes listed in Table 10 on page 58, the following
messages may be encountered:
Table 10: Error Codes