Power-On Settings and Service Functions
LIGHT
PARAM
48
•
Data Port Mode (ASCII, OXINET, CLINICAL, GRAPH, AGILENT
[Agilent HP monitor], SPACELB [SpaceLabs monitor], MARQ [GE
Marquette monitor], DATEX [Datex-Ohmeda AS/3 monitor]). Available
selections depend on the software installed in your N-595.
•
SpO
Upper Alarm Limit (lower limit plus 1 to 100%)
2
•
SpO
Lower Limit (80% to upper limit minus 1)
2
Use the following procedure to set institutional defaults.
1. Disconnect the O
M
XI
Note:
If the O
M
sensor is not disconnected, the only softkeys on the screen
XI
AX
will be PARAM and EXIT.
2. Set desired parameters to the institutional values. Refer to the N-595
Operator's Manual for the procedures to set the values.
3. Simultaneously press the LIGHT softkey and the CONTRAST button until
the menu bar changes to the softkey headings shown below.
4. Press the PARAM softkey.
sensor from monitor.
AX