Nellcor OxiMax N-595 Service Manual page 40

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

Performance Verification
Test #4: Light
32
8. The monitor will display:
75 %SpO
2
60 BPM
alarm
pulse amplitude indicator - low level modulation
9. Perform Test #1: BPM on page 28. The pulse amplitue indicator should
indicate low level modulation.
10. Perform Test #2: SpO2 on page 29. The pulse amplitue indicator should
indicate low level modulation.
1. Press the SRC-MAX Light Level selection button. The SRC-MAX Light
Level
LED will light.
2. The monitor pulse amplitude waveform will initially increase in amplitude
and then stabilize.
3. The monitor will display:
75 %SpO
(test pass criteria is 73 to 77 %SpO
2
60 BPM (test pass criteria is 57 to 63 BPM inclusive)
alarm
pulse amplitue indicator - high level modulation
inclusive)
2