3.5

Set the Pressure Module to Zero

Use this option to write a new zero pressure value to the
pressure module in use. The sensor adjustment is permitted if
it obeys the condition that follows:
• Adjustment ≤10% FS positive pressure value (for the
Sensor)
Note: To make a temporary adjustment for zero, use the Tare
function
3.6

Example Procedure: Measure Pressure

1
Assemble the pressure indicator with the correct PM
620 modules
2
Set the applicable software options
06 June 2013
Druck DPI620 Genii
Page 73 of 179