Direction 5936427, Rev. 1
Section 6-4

DICOM Server Service Adjustments

6-4-1

Slow response rate between the LOGIQ Totus and a DICOM (PACS) server

Problems with slow responses may result in images being re-sent automatically and low transfer rates.
If you are experiencing problems with slow responses from a DICOM server, increase the timeout
setting.
To adjust properties for a DICOM server (online):
1.) Navigate to Utility > Connectivity > Dicom > Service.
2.) Select the service for the DICOM server.
3.) Under Connection, set Timeout (sec) to 360.
6-4-2

Shutdown problems when running a portable LOGIQ Totus

When portable (offline and not communicating with a PACS server), use the following settings to avoid
problems with shutdown speed:
To adjust properties for a DICOM server (offline):
1.) Navigate to Utility > Connectivity > Dicom > Service.
2.) Select the service for the DICOM server.
3.) Under Connection, set these parameters:
-
-
-
Section 6-5

Volume Navigation (VNAV) Calibration

There is no calibration available in the field. If the registration ms to be out of calibration, reloading the
software will restore the factory settings. If the registration still ms out of calibration, replace the Volume
Navigation (VNAV) hardware.
For information on the hardware replacement,
on page 8-210
6 - 14
GE HealthCare CONFIDENTIAL
Maximum Retries - Set to 0.
Retry Interval (sec) - Set to the lowest possible value.
Timeout (sec) - Set to the lowest possible value.
.
Section 6-4 - DICOM Server Service Adjustments
LOGIQ Totus Basic Service Manual
8-11-19 "Replacement of the VNAV Option Cable Set-U"