Nikon Eclipse TI-TIRF-E Setup Manual page 94

Illuminator unit/motorized illuminator unit/photo activation illuminator unit, beamsplitter 50/50
Table of Contents

Advertisement

6
Attach the laser safety cover (TIRF illuminator unit or motorized TIRF illuminator unit only).
Safety interlock (both sides)
Figure 4.2-12
7
Attach the light shielding plate.
Figure 4.2-13
Chapter 4 Setting Up the Microscope
4.2 Setup Procedure
Laser safety cover
ON
OFF
6V30W
12V100W
MAX.
Upper cover
Body
Light shielding plate
If you are using the TIRF illuminator unit or the
motorized TIRF illuminator unit, the motorized shutter
will not open unless the laser safety cover is attached
on the stage. Make sure it is attached in the prescribed
position.
1.
Place the laser safety cover body on the stage,
and secure it by tightening the 4 screws with the
Allen wrench.
2.
Connect the laser safety cover's cable to the
safety-cover connector of the TI-LUSU shutter
unit.
3.
Attach the cover on top of the body.
The interlock will close the motorized shutter of
the laser unit if the cover is removed during use.
* In the case of the photo activation illuminator unit,
you must attach a specimen cover on the stage
when using episcopic illumination or laser light.
* Touching the interlocking part or getting it wet could
cause it to malfunction.
Using an Allen wrench, screw the light shielding plate
onto the eyepiece tube.
1-80