Nikon Eclipse TI-TIRF-E Setup Manual page 126

Illuminator unit/motorized illuminator unit/photo activation illuminator unit, beamsplitter 50/50
Table of Contents

Advertisement

Polarizer-equipped tool
(Push the tool until its
edge surface touches
the dichroic mirror.)
Adjusting screw
on the incoming
laser side
Adjusting screw
on the fiber
Coupler
Target (such as paper)
Direction of
polarization
of polarizer
11
Laser intensity measurement
At first, maximize each laser intensity. Move off all of movable ND filters from the optical path and
adjust Argon laser to its maximum power. Measure and record the light intensity of each laser emitting
from the coupler with a calibrated laser intensity power meter.
This completes the rough setup of the lasers.
Chapter 5 Laser Unit
5.2 C-LU2 Two-laser Unit, and C-LU3 Three-laser Unit
L1
A
R1
B
Rotate polarizer so that the argon
laser becomes darker.
Direction of polarization of laser
R1
L1
Figure 5.2-13
1-112
He-Ne2
Ar
He-Ne1

Hide quick links:

Advertisement

Table of Contents
loading