Nikon Eclipse TI-TIRF-E Setup Manual page 234

Illuminator unit/motorized illuminator unit/photo activation illuminator unit, beamsplitter 50/50
Table of Contents

Advertisement

Chapter 6 Connection between the Laser Unit and the Microscope
2
r command: Data request command
a: Normal
r command
n: Error response
rTSR
rTTS(1)
rTES
rTSW
rVER
rPNM
6.1 Setting and Connecting the TI-LUSU Shutter Unit
response
The command reads the shutter condition.
(1): [0] means all shutters are closed. [1] to [3] means the shutter of the
aTSR(1)
number is open.
nTSR(6)
(6): Error code
* The rTSR command is disabled in the shutter independent control
mode.
The command reads the shutter open time setting.
aTTS(2)
(1): [1-3] Shutter number
nTTS(6)
(2): [0000-1000] Time setting
(6): Error code
The command reads the open/close conditions of the eyepiece tube
optical path and the safety cover.
(3): [1] means that either of the binocular route or the safety cover is
aTES(3)
open. [0] means both are closed.
nTES(6)
(6): Error code
* The safety cover open/close condition is detected only for the
TIRF/motorized TIRF illuminator system.
The command reads the setting for the laser open/close switch on the
aTSW(4)
control panel.
nTSW(6)
(4): [1] Enabled, [0] Disabled
(6): Error code
The command reads the program version.
aVER(5)
(5): V*.** (Five characters. "*" is a numeral. Example: V1.00)
nVER(6)
(6): Error code
The command reads the program name.
aPNM(5)
(5): T-LUSU
nPNM(6)
(6): Error code
Table 6.1-4
Description