Nikon Eclipse TI-TIRF-E Setup Manual page 207

Illuminator unit/motorized illuminator unit/photo activation illuminator unit, beamsplitter 50/50
Table of Contents

Advertisement

5
Fixing the beam shift part and the optical fiber coupling device
Rotate by hand
Horizontal clamp screw
Figure 5.4-103
Figure 5.4-104
Chapter 5 Laser Unit
5.4 LU4A Four-laser Module A
Vertical clamp screw
(nearer to the axis A)
Axis A
Knurled screws
1.
Tighten the clamp screws for the horizontal direction
and the vertical direction to fix the L3 beam shift part.
2.
Tighten the center knurled screw of the optical fiber
coupling device. And then, gradually tighten two
other knurled screws to fix the optical fiber coupling
device.
1-193