WARNING
Use of controls or adjustments or performance of procedures other than those specified in this manual
may result in hazardous radiation exposure.
1.
Laser class
This product is a laser apparatus that uses Class 3B lasers (air-cooled Ar lasers, He-Ne lasers,
semiconductor lasers). Accidental exposure of eyes or skin to laser light may result in injury or
other problems. Confirm that there is no one in the area before turning on the lasers. Always wear
laser-protection goggles when using the laser.
2.
Laser safety
This product is designed and manufactured in compliance with the Performance Standards for
Light-Emitting Products established by the U.S. FDA and the IEC Laser Product Safety Standard
(IEC 60825-1), except for deviations pursuant to Laser Notice No.50. Those using this laser
product must complete laser safety courses and setup training. Users are requested to take all
appropriate safety measures specified by the preceding standards, in accordance with local laws
and regulations.
The laser safety officer (LSO) must take charge of this laser product and users should follow the
instructions from the LSO.
FDA Class 3B Laser Product (Laser Notice No. 50)
IEC Class 3B Laser Product
Wavelength: 400 to 700 nm (CW)
Beam divergence: 0.01 to 2.91 rad
Maximum power or energy output: total power 500 mW Max. (CW)
3.
Manuals and accessories
All manuals and accessories used in this laser product and provided by the distributor must be
provided to the user.
4.
Intended application of the product
This product can be used only for microscopic observations. This product is required to be
attached to the specific Nikon microscope, Ti-E, Ti-E/B, Ti-U, Ti-U/B. Do not use this product for
other purpose.
5.
Setting up the product
Setup of this product must be handled by trained service personnel.
This task should not be performed by anyone other than the trained service personnel. If
performed, it may result in improper setup, impairing system performance and exposing the eyes
or skin to the laser light, resulting in injury.
Safety Precautions
LASER RADIATION
AVOID EXPOSURE TO BEAM
CLASS 3B LASER PRODUCT
4