(3) EMISSION key switch
The switch turns on lasers and emits lasers.
The key is in vertical position ( O ):
The key is in horizontal position ( I ): The lasers start emitting.
(4) Power switch
This is the main switch of the LU-LR four-laser PS rack.
When turning on this switch only, the laser does not start firing.
(5) INTERLOCK connector
The connector is used to disable the remote interlock function. When the connector is open, no laser
emits (remote interlock status). To disable the remote interlock function, attach the provided interlock
jumper connector.
(6) Cooling fan
The fan cools the power supply devices in the LU-LR four-laser PS rack. When lasers are energized, the
fan starts.
(7) Power breaker for 15 A line
The breaker protects over-current of the LU-LR four-laser PS rack controller and the Ar laser power
supply. Normally it is turned on.
(8) Power breaker for 5 A line
The breaker protects over-current of the lasers other than Ar laser. Normally it is turned on.
(9) Caution label
(10) Name plate for 15 A line ratings
(11) Name plate for 5 A line ratings
(12) AC inlet for 15 A line
The inlet supplies power to the LU-LR four-laser PS rack controller and the Ar laser.
When no power is supplied to the inlet, the four-laser PS rack does not work.
CAUTION
Use a power cable that satisfies the power ratings of the AC inlet and the safety standard of
the country.
(13) AC inlet for 5 A line
The inlet supplies power to lasers other than the Ar laser.
CAUTION
Use a power cable that satisfies the power ratings of the AC inlet and the safety standard of
the country.
Chapter 5 Laser Unit
5.1 Overview of the Laser Units
The lasers stop emitting and the key can be extracted.