Illuminator unit/motorized illuminator unit/photo activation illuminator unit, beamsplitter 50/50
Table of Contents

Advertisement

DSC support columns (2 pieces)
Attach supports to fix the camera to the anti-vibration table.
DSC support columns
Figure 4.2-22
STORM slider
Insert the dedicated optical system into the TIRF optical system to perform N-STORM microscopy.
OUT
IN
STORM slider
Figure 4.2-23
Lambda plate slider
Insert an optical device for polarization adjustment to perform N-STORM microscopy.
Lambda
plate slider
Figure 4.2-24
Chapter 4 Setting Up the Microscope
4.2 Setup Procedure
Screw the supports in the screw holes in the camera
(Andor iXon) body.
Secure the camera to the side 3D-STORM port of the
microscope and set the magnet base to ON.
Detach the cover at the upper part of the laser
introduction section in the motorized TIRF illuminator
unit.
You can see a slot to insert the STORM slider. Insert
the slider into the slot.
Then attach the cover to prevent the slider from
coming off.
Insert this slider during N-STORM microscopy, but
slide it out of the optical path in other cases.
Remove the stopper screw and detach the ND16 of
the motorized TIRF illuminator unit.
Insert the lambda plate slider into the same position.
Install the stopper screw.
Insert this slider during N-STORM microscopy, but
slide it out of the optical path in other cases.
1-86