Figure 5.3-44
Figure 5.3-45
This completes the adjustment of the Ar dichroic mirror adjustment unit.
Chapter 5 Laser Unit
5.3 C-LU3EX Three-laser Unit EX
5.
Adjust the vertical axis setscrew and align the
vertical axis of the laser light with the tool opening
position.
6.
Carefully tighten the horizontal clamp screws.
7.
If the beam becomes misaligned from the center
after tightening, fine-adjust the vertical setscrew to
align the beam with the tool opening.
8.
Carefully tighten the vertical clamp screw.
1-141