Nikon Eclipse TI-TIRF-E Setup Manual page 320

Illuminator unit/motorized illuminator unit/photo activation illuminator unit, beamsplitter 50/50
Table of Contents

Advertisement

Chapter 6 Connection between the Laser Unit and the Microscope
Holder plate locking screws
Figure 6.3-4
CAUTION
To using the laser beam for passing through the mirror, if the mirror is not set in a collimated angle,
interference fringes may appear in the PA-GFP illumination. Carefully set the mirror in a collimated
angle except when using a total reflection mirror.
(Standard collimated angle for the half mirror is 5".)
6.3 Replacement of Optical Path Switch Mirror
4.
Remove the two holder plate locking screws. And then,
remove the holder plate and flat spring to replace the
mirrors.
Mirror size: 36 x 25.7 x 1
Holder plate
5.
Assembly is the reverse of disassembly.
Place the reflection surface of the mirror downward into
Flat spring
the cutout of the mirror frame.
Mirror
There is a positioning hole on the mirror unit. Securely
(Place the
attach the mirror unit to the illumination attachment to fit
reflection
the pin on the main body.
surface
downward.)
2-62