Nikon Eclipse TI-TIRF-E Setup Manual page 62

Illuminator unit/motorized illuminator unit/photo activation illuminator unit, beamsplitter 50/50
Table of Contents

Advertisement

TI-FLC-E motorized epi-fl filter turret or TI-FLC-E/HQ motor epi-fl filter turret
Filter cube port
(and its cover)
Front
TI-LUSU shutter unit
Safety cover signal connector
Laser unit connector
Serial IF connector
RS-232C
LASER SHUTTER
Laser shutter
switch / indicator
Exclusive use for TI-TIRF system
Don't use this unit with other systems.
TI-LU4SU shutter unit LU4
Laser unit connector
TI-LU4SU
Option switch
Laser shutter
switch
Chapter 1 System Configurations and Part Names
1.4 Name of Each Part
TI-FLC-E motorized epi-fl
filter turret or TI-FLC-E/HQ
motorized epi-fl filter turret
Shutter lever (O: open, C: close)
Figure 1.4-18
Eyepiece optical path
signal connector
AC adapter
connector
INPUT
LASER UNIT
12V
SAFETY
1A
COVER
CLOSE
BINO
CLOSE
POWER
TI-LUSU
Figure 1.4-19
Indicator lamp
LASER
EMISSION
POWER
Power lamp
Figure 1.4-20
The epi-fl filter turret is used to select an excitation
method by rotating the turret to change filter cubes
including a dichroic mirror.
The shutter unit controls motorized shutters in the
laser unit.
Use this unit in combination with the C-LU2, C-LU3 or
C-LU3EX laser unit.
Safety cover
close lamp
Eyepiece optical
path closed lamp
Power switch
Power lamp
The shutter unit controls motorized shutters in the
laser unit.
Use this unit in combination with the LU4A laser unit.
If the safety interlock functions, all the shutters are
closed, and the lamp turns off.
To unlock the safety interlock, solve what caused the
safety interlock being locked, then cancel the safety
interlock from the software.
1-48

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents