Laser Rough Adjustment - Nikon Eclipse TI-TIRF-E Setup Manual

Illuminator unit/motorized illuminator unit/photo activation illuminator unit, beamsplitter 50/50
Table of Contents

Advertisement

5.2.3

Laser Rough Adjustment

Determine the position of the argon laser by using the optical axis of the farthest He-Ne laser (He-Ne1 laser in
Figure 5.2-3 or in Figure 5.2-4) as the reference. Align the lasers so that their laser beams can be transmitted
through the optical fiber for the excitation light.
1
Setting up the dichroic mirror centering tool
Remove the dichroic mirrors and mount the dichroic mirror centering tool (see Figure 5.2-5). Install the
lasers so that their incident points are on the tool.
Remove the
dichroic mirrors
Remove
the dichroic
mirrors.
Pinhole for mounting
the dichroic mirror
centering tool
After the dichroic mirror of the He-Ne2 laser
has been removed
Chapter 5 Laser Unit
5.2 C-LU2 Two-laser Unit, and C-LU3 Three-laser Unit
Figure 5.2-3
Figure 5.2-4
Dichroic mirror centering tool
Figure 5.2-5
1-106
He-Ne1
Ar
He-Ne2
Ar
He-Ne1