Ower Sensor Calibration Factor Entry - HP 8753D Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753D Option 011:
Table of Contents

Advertisement

Power Sensor Calibration Factor Entry
5. Zero and calibrate the power meter and sensor.
6. ress
ERVICE MENU TE T OPTION
SYSTEM 5
CAL FACTOR EN OR A
A (H 8482A).
7. Build a table of up to 12 points (12 frequencies with their calibration factors). To enter
each point, follow these steps:
a. ress
ADD FREQUENCY
b. Input a frequency value and then press the appropriate key: (
c. ress
CAL FACTOR
frequency you entered.
The cal factor and frequency values are found on the back of the sensor. If you make a
mistake, press
5
d. ress
to complete the data entry for each point.
DONE
ote
The following terms are part of the sensor calibration menu:
EGMENT
EDIT
DELETE
ADD
CLEAR LI T
DONE
3-10 Adjustments and Correction Constants
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
to access the calibration factor menu for power sensor
.
and enter the calibration factor percentage that corresponds to the
and re-enter the correct value.
Allows you to select a frequency point.
Allows you to edit or change a previously entered value.
Allows you to delete a point from the sensor cal factor table.
Allows you to add a point into the sensor cal factor table.
Allows you to erase the entire sensor cal factor table.
Allows you to complete the points entry of the sensor cal
factor table.
LO / EN R LI T
,