Recorder Output; Test Setup For Recording Swept Measurements - HP E4419B User Manual

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

Recorder Outputs
The rear panel Recorder Output connectors (A and B) produce a dc voltage
that corresponds to the power level in Watts of the channel, depending on
the measurement mode. This dc voltage ranges from 0 to +1 Vdc. The
output impedance is typically 1 kΩ . Channel and display offsets, and duty
cycle have no effect on the Recorder Outputs.
For example, the Recorder Outputs can be used to;
record swept measurements on an X-Y recorder
level an output from a source using external leveling, or
monitor the output power on a strip chart recorder. A setup for
recording swept measurements is shown in Figure 2-20.
See Figure 2-22 for details on which power meter functions are
implemented in the recorder output.
Figure 2-21: Test Setup for Recording Swept Measurements
Swept Source
SWEEP
RF OUT
OUT
X-AXIS (FREQUENCY)
To access the "Recorder" menu press
,

Recorder Output

channel you want to set up. This menu allows you to switch the Recorder
Output signal either on or off. The
allow you to enter the input power level that you want to represent the
1 V
maximum and 0 V
dc
Output.
HP E4419B User's Guide
Get other manuals https://www.bkmanuals.com
Device
Under Test
IN
OUT
X-Y Recorder
System
Inputs
or
Channel A
Channel B
Max Power
minimum output voltage of the Recorder
dc
Power Meter Operation
Recorder Outputs
RECORDER
OUTPUT A
Power Meter
CHANNEL A
INPUT
Y-AXIS (POWER)
,
,
More
depending on which
and
Min Power
softkeys
2-65

Advertisement

Table of Contents
loading

Table of Contents