Section5

Setting Output Processing of Sensing Information

Setting Output Processing of Sensing Information

Set how sensing information is to be processed for outputting the required values.

Setting the Scaling

This setting is used when you want to correct any errors that are generated due to the
installation status of the Sensor Head, and display the corrected value on the main display.
Place an actual sensing object in position for measurement.
There are three setting modes: "manual setting of correction values," and "one-point
scaling" and "two-point scaling" that automatically set the correction values of a placed
sensing object.
When GLASS mode is selected for the measurement object, the menu used only for
scaling is displayed.
Measuring the Thickness of Transparent Objects p.3-4
The settings listed below return to the default settings when scaling is set. Set these items after scaling
settings have been completed.
• Zero reset
■ Setting One-point Scaling
Measurement is performed at one position and offset values are set for that measure-
ment value.
The offset and increment/decrement inversion (display inversion) can be set.
Example: Displaying the height of the sensing object
Measuring center
distance
Measured value
-4.5 mm
ZS-HL
5-14
User's Manual
Sensor Head
Target display value
Sensing
10 mm
object
Display value
(mm)
10
-4.5
Measured value
(mm)
-4.5