Omron ZS-HL User Manual page 129

Smart sensor 2d cmos laser type
Hide thumbs Also See for ZS-HL:
Table of Contents

Advertisement

● When Measured Value is Output Continuously
ON
Laser beam emission
OFF
Internal processing
Judgment output
OFF
GATE signal
ON
Terminal block output
Linear output
Item
T1
Time during which the mea-
sured value is determined
from the laser beam emis-
sion
T9
Update cycle of terminal
block output
T10 OFF time of GATE signal
T18 Update cycle of terminal
block output 2
T21 Response time of GATE
signal
T11 Response time of linear
output
To the terminal block, only the result of the TASK for which the measurement is performed is output.
You can judge which TASK number it is by checking the lower 2 bits of the 16-bit output value.
To the linear output, only one selected TASK result is output.
T1
T21
T10
TASK2
TASK3
TASK4
TASK1
T11
TASK *
TASK *
TASK *
TASK *
Minimum
Average
-
Sampling cycle × update cycle
68 µs
Sampling cycle × update cycle × (number of
TASKs that have been performed)
320 µs
Timing Charts
T18
T9
TASK2
TASK3
TASK4
TASK *
TASK *
TASK *
Maximum
Sampling cycle ×
3
500 µs
Section6
TASK1
TASK *
Remarks
The update cycle
can be set in [I/
OSET]-[TERMI-
NAL BLOCK]-
[CYCLE].
When the mea-
surement is per-
formed
continuously and
the amount of light
is stable
It changes
depending on the
way the measured
value fluctuates.
ZS-HL
6-25
User's Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents