Omron ZS-HL User Manual page 125

Smart sensor 2d cmos laser type
Hide thumbs Also See for ZS-HL:
Table of Contents

Advertisement

■ PEAK/BOTTOM/P-P/AVERAGE HOLD
This is the timing charts when the trigger mode is external trigger.
The result processed at the timing drawn in diagonal line is the target of hold. For the
sampling period, see the settings for [HOLD].
Setting HOLD Functions p.5-18
OFF
Trigger input
ON
OFF
Trigger rest input
ON
Internal processing
Judgment output
OFF
BUSY output
ON
Terminal block output
OFF
GATE signal
ON
Linear output
Item
T2
Minimum time of trigger
input
T12 Timing of starting and end-
ing the measurement after
trigger input
T13 Response time of judgment
output, BUSY output, and
terminal block output
T10 OFF time of GATE signal
T21 Response time of GATE
signal
T11 Response time of linear
output
T2
T12
T12
T13
T13
Sampling period
T21
Aver-
Minimum
age
Sampling cycle
× 4
Sampling cycle
× 3
Sampling cycle
× 4
68 µs
320 µs
Timing Charts
T2
T13
T10
T21
T10
T11
T11
Maximum
Sampling cycle × 4
Sampling cycle × 5 When the terminal
block output is in the
update cycle 1
500 µs
It changes depend-
ing on the way the
measured value fluc-
tuates.
Section6
Remarks
ZS-HL
6-21
User's Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents