High Calibration Connection Diagrams - Mallinckrodt INOmax DSIR Pocket Manual

Delivery system, series 3 software
Hide thumbs Also See for INOmax DSIR:
Table of Contents

Advertisement

High Calibration
Connection Diagrams
Connection Diagram for NO and NO
Range Calibration
1. Cylinder Cap
2. Regulator Seal
3. Regulator
4. Tubing Adapter
5. Attach Tube Kit
6. Turn valve counter-clockwise to
start gas flow
Part No. 20751 Rev-01
2014-08
6a
6
6a. If the pressure is in the red or black
zone (0-25 psig) select another INOcal
cylinder.
7. Attach tube kit to sample line
8. Calibrate sensor
9. Remove sample line from tube kit
10. T urn valve clockwise to stop gas flow
11. One-way Valve
56
High
2
7
To
INOmax DS