High Calibration
Connection Diagrams
Connection Diagram for NO and NO
Range Calibration
1. Cylinder Cap
2. Regulator Seal
3. Regulator
4. Tubing Adapter
5. Attach Tube Kit
6. Turn valve counter-clockwise to
Part No. 20751 Rev-01
2014-08
High
2
7
6a
6
6a. If the pressure is in the red or black
zone (0-25 psig) select another INOcal
cylinder.
7. Attach tube kit to sample line
8. Calibrate sensor
9. Remove sample line from tube kit
11. One-way Valve
56
To
INOmax DS
IR