Objet30 User Guide
4. Select the confirmation check box, and click Next.
5. In the next wizard screen, observe the numbers and wait until the weight
6. Select the check box, and click Next.
DOC-34025 Rev. I
Figure 6-100 Confirmation of cartridge removal
level is relatively stable—two units above or below the average level
shown.
Figure 6-101 Weight stabilization
If you need to calibrate more than one load cell, the next calibration screen
appears. Repeat steps 5 and 6 until all of the load cells are calibrated.
6 Operating and Maintaining the Printer
6-76