BD FocalPoint™ GS Imaging System Instrument User's Manual
The following table lists the Cautions and Warnings found in this manual.
Notice
Located on
Type
Warning
Page 23
Caution
Page 32
Caution
Page 50
Caution
Page 60
Caution
Page 60
Warning
Page 137
Warning
Page 138
Warning
Page 138
Warning
Page 141
Caution
Page 141
Caution
Page 142
Caution
Page 153
202
Found on
Notice
BD FocalPoint™
Make sure that the clip grips the slide edge securely. An unsecured
GS Review
slide could fall into a slotted stage opening and be damaged.
Station
Damaged or broken slides can cause injuries. Be careful not to allow
the clip to snap against the glass.
BD FocalPoint™
All FOVs must be reviewed before performing full slide review.
GS Review
Station
BD FocalPoint™
This feature provides the history of BD FocalPoint™ GS Review
GS Review
Station slide information only, not the patient history.
Station
BD FocalPoint™
When the server is configured for automatic data transfer, do not
GS Review
perform a manual data transfer.
Station
BD FocalPoint™
Your images are stored on the BD FocalPoint™ Slide Profiler for 10
GS Review
days. Your copy service settings should be set to less than 10 day
Station
intervals.
BD FocalPoint™
Shock and fire potential from spilled liquids. Do not place containers
Slide Profiler
with liquids on the device or the workstation cart. Do not spill liquids
on the system; fluid seepage into internal components creates a
potential shock and fire hazard. Shut down the device, disconnect
from the power source and wipe up all spills immediately. Do not
operate the system if internal components have been exposed to
fluid.
BD FocalPoint™
Be careful not to drop or break slides during slide preparation and
Slide Profiler
when loading and unloading slides into trays. Broken slides may
cause injuries.
BD FocalPoint™
Electromagnetic fields. This is a Class A product. In a domestic
Slide Profiler
environment, this product may cause radio interference with other
electronic devices, such as telephones and other medical equipment,
in which case the user may be required to take measures to reduce
such interference.
BD FocalPoint™
Remove all obstructive jewelry before loading or unloading trays.
Slide Profiler
After opening a hopper door, be sure all moving parts in the hopper
have stopped before inserting or removing a tray. If trays are inserted
before all moving parts have stopped, injury may result or the
instrument may malfunction.
BD FocalPoint™
If slides that do not match the operating mode of the instrument are
Slide Profiler
successfully processed on the instrument, the slides results reporting
for the print set will be invalid. All the slides in the print set must be
rerun on an instrument with the correct operating mode.
BD FocalPoint™
When adding trays to the input hopper during processing, be careful
Slide Profiler
not to touch the extended edge of a tray currently being processed.
Touching the tray could invalidate the positioning determinations
made for the cellular features on the slide in process.
BD FocalPoint™
Contact BD Technical Support before attempting to shut down the
Slide Profiler
instrument.
Follow the shutdown procedure exactly.