Chapter 3: Testing Logic Analyzer Performance
Prepare the Logic Analysis System for Testing

Perform System Self-Tests

5 Do a self-test on the logic analysis system.
a When the logic analysis system has finished booting, the Waveform
window appears. Select Help→Self-Test... from the main menu. The
Analysis System Self Tests window will appear.
b In the Select Suite(s) list, select <all>. This will cause <all> to be
selected in the Select Test(s) list.
c Select Start. This will perform a complete system self-test.
The progress of the self tests is displayed in the Progress & Statistics area of the
window.
d When the self-tests are complete, check the Results window to ensure
that the Result Summary says that all tests passed. If all tests did not
pass, refer to "Troubleshooting" on page 63.
e Select the Close button to close the Analysis System Self Tests window.
f If all module self-tests pass, then record "PASS" in the "Logic Analysis
System Self-Tests" section of the Performance Test Record (page 60).