To perform self tests

Use this test when one of the following occurs:
Performance test fails
Calibration fails
Trigger fails
GPIB fails
Microphone power fails
Serial port fails
Parallel port fails
Failure is intermittent
Step 1. Run all the functional self tests.
• Connect the rear panel SOURCE output to the rear panel TACH input
using a BNC cable.
• Remove all cables from the front panel input connectors.
The ICP self test outputs approximately 30 Vdc on the input connectors. Before
starting the self tests, disconnect all devices connected to the input connectors.
Devices left connected during the ICP self test may be damaged.
• Set the power switch to on (
• After calibration is complete, press the following keys:
[
]
Input
[
ALL CHANNELS
[
CH* FIXED RANGE
[
]
System Utility
[
]
CALIBRATN
[
OFF ]
AUTO CAL
[
]
Rtn
[
]
MORE
[
]
SELF TEST
[
FUNCTIONL TESTS
[
]
ALL
[
]
CONTINUE
).
l
]
]
]
Troubleshooting the Analyzer
To perform self tests
4-31