Specifications (oscilloscope option)

The specifications are the performance standards against which the
1670G-series logic analyzers oscilloscope is tested.
Bandwidth*:
Time Interval Measurement
Accuracy*, **:
DC Offset Accuracy*:
Voltage Measurement
Accuracy*:
Trigger Sensitivity*:
dc to 50 MHz:
50 MHz to 500 MHz:
Input R:
* = Specifications valid within +/-10 °C of self-calibration temperature
** Specification applies at the maximum sample rate. At lower rates,
specification should be +/-(0.005% x ∆ t) + (2 x 10-6 x delay setting) +
(0.15 X sample interval) for bandwidth limited signals (tr = 1.4 x
sample interval). Sample interval is defined as
dc to 500 MHz (realtime, dc coupled)
+/-[(0.005% of ∆ t) +
-6
(2 x 10
* delay setting) + 150 ps]
+/-(1.0% of channel offset + 2.0% of
full scale)
+/-((1.5% of full scale +
offset accuracy + 0.016 V/div)
+ 0.008 * V/div )
0.063 * full scale
0.125 * full scale
1 M Ω +/-1%,50 Ω +/-1%
---------------------------- -