DO output
4.9.5.1. FULL SCALE 2 output
Description
● Select a contact output as DO1 and/or DO2 at FULL SCALE2 measurement status.
● RANGE FULL SCALE 2 contact output is activated for either the volume flow rate or mass flow rate set with the analog output source
designation (see section 4.8.1.3).
FULL SCALE 2
: Contact output is activated during FULL SCALE2 measurement. (AUTO 2, BI-DIR, BI-DIR AUTO2)
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
Set the DO1 output to "FULL SCALE2".
(example)
Also, set the contact to "ACTIVE ON".
Key operation
Press the
▼
Press the
ENT
▼
Press the
▼
* Use the
Press the
ENT
▼
Press the
▼
Press the
ENT
▼
Press the
ENT
▼
Press the
ENT
▼
* To select normally on, press the
Press the
ESC
the measurement mode.
Description
key twice to display "OUTPUT SETUP".
key once to display "ZERO ADJUSTMENT".
ENT
key for 7 times to display "DO1 OUT".
key to select "DO2 OUT".
key once to blink the cursor.
ENT
key for 3 times to display "FULL SCALE2" on the 2nd line.
key once to register "FULL SCALE2".
ENT
――― "FULL SCALE2" has been registered. ―――
key once to display "CONTACT ACTION".
ENT
key once to register "ACTIVE ON"(normally off).
ENT
――― ACTIVE ON has been registered. ―――
key twice and then press the
ESC
key.
key for 3 times to enter
-66-
Display
OUTPUT SETUP
ZERO ADJUSTMENT
DO1 OUT
NOT USED
DO1 OUT
NOT USED
DO1 OUT
FULL SCALE2
DO1 OUT
COMPLETE
↓
STATUS OUT
CONTACT ACTION
CONTACT ACTION
ACTIVE ON
CONTACT ACTION
COMPLETE
↓
STATUS OUT
CONTACT ACTION
0.000
%
0.000
m3
CLEAR