Calibrating The Measured Value - Fuji Electric FSJ Instruction Manual

Ultrasonic flowmeter for steam
Table of Contents

Advertisement

Calibrating the measured value

Description
● Measured value can be calibrated manually.
Zero point and span adjustment can be made.
Settable range
(1)
Zero point
: -5 to +5 [m/s] in terms of flow velocity
in piping.
(2)
Span
: ±200%
The output value (reading, analog output and total output)
is computed by the following expression.
Measured value × [Span set value %]
Output =
<Note> The flow rate unit is the unit selected at "VOLUME FLOW UNIT".
Specify the "VOLUME FLOW UNIT" setting before setting. (See Section 4.7.2)
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
Compensate the zero point to 0.5m3/h, and the span by +1%.
(example)
* Set the PIPE PARAMETER and VOLUME FLOW UNIT beforehand.
Key operation
Press the
Press the
ENT
Press the
Press the
ENT
Press the
Press the
Press the
ENT
――― "CALIBRATION ZERO" has been registered. ―――
Press the
Press the
ENT
Press the
Press the
Press the
ENT
――― "CALIBRATION SPAN" has been registered. ―――
Press the
ESC
the measurement mode.
+ Zero point
100
Description
key twice to display "OUTPUT SETUP".
key once to display "ZERO ADJUSTMENT".
ENT
key for 9 times to display "CALIBRATION ZERO".
key once to blink the cursor.
ENT
key for 6 times to move the cursor.
key for 5 times to set "5".
key once to register.
ENT
key once to display "CALIBRATION SPAN".
key once to blink the cursor.
ENT
key twice to move the cursor.
key once to set "1".
key once to register.
ENT
key once and then press the
ESC
key for 3 times to enter
-75-
Display
OUTPUT SETUP
ZERO ADJUSTMENT
CALIBRATION ZERO
0.000 m3/h
CALIBRATION ZERO
00000.000 m3/h
CALIBRATION ZERO
00000.000 m3/h
CALIBRATION ZERO
00000.500 m3/h
CALIBRATION ZERO
 COMPLETE 
CALIBRATION ZERO
0.500 m3/h
CALIBRATION SPAN
100.0 %
CALIBRATION SPAN
100.0 %
CALIBRATION SPAN
100.0 %
CALIBRATION SPAN
101.0 %
CALIBRATION SPAN
 COMPLETE 
CALIBRATION SPAN
101.0 %