(1) Analog output
• Use supplied dedicated cable (DB-25P 25 male pin) for taking out output signals.
Fasten the cable securely using supplied screws.
Output signal:
Permissible load: 4 to 20mA DC, 550Ω or lower
• The analog output corresponds to the channel (Ch) on the measurement screen display.
CAUTION
None of the analog outputs of this product is isolated. To prevent the effect of external interference,
individually isolate signals before drawing out wiring to outdoors or extending the dedicated cable.
(2) Communication output
• Digital output signals are RS232-C Modbus capable.
Individual programming by referring to the attached "Description of Communication Function" is
required.
Use a commercially available cable (D-sub 9 pin, female-female).
INZ-TN2ZSVS-E
4 to 20mA DC or 0 to 1V DC (by code symbols), non-isolated
0 to 1V DC, 100kΩ or higher