3.10.1. DO output settings

Description
Selects the total pulse and status (warning and flow rate switch, total switch, etc.) output.
DO output types (common to DO1 and DO2):
Settable range
Not used
All alarm
Device error
Process error
Range-over/Pulse range-over :
Negative flow direction
Forward total flow pulse
<Note> Only DO1 is set for forward total flow pulse.
Contact operation
Operation ON
Operation OFF
When the contact operation setting is "Operation OFF", DO is output when the power is turned ON.
Before setting, confirm in advance that you want to change the DO output.
<Note> DO output specification
DO1/DO2
For specific key operations, refer to the example in

3.10. Status settings

:
: Does not use the contact output.
:
: Contact output is activated upon instances of device error or process error
:
: Contact output is activated when a circuit error (memory, etc.) or a temperature circuit
error occurs.
:
: Contact output is activated when no waves are received or waves are unstable.
: Contact output is activated when the instantaneous flow rate exceeds the upper limit
of 120% or lower limit of -20% of the range.
Contact output is activated when the flow rate total pulse output exceeds the maximum
frequency limit.
:
: Contact output is activated when the flow is in the reverse direction.
:
: Outputs the forward total flow pulse.
:
: Normally OFF
:
: Normally ON
Open collector, contact capacity 27.5 V DC, 100 mA
:
When total pulse output is selected
(Note: Refer to
3.10.2
100 pulses/sec or less (at full scale flow rate)
Pulse width: 5 ms, 10 ms, 50 ms, 100 ms, 200 ms
0.
Total flow pulse setting (total rate, pulse width))
- 27 -