4.8.2.3. Setting the MASS TOTAL pulse (MASS TOTAL rate, pulse width)

Description
● If integrating mass into the integrator when measuring mass flow rate, specify the output pulse as follows.
Mass can be integrated at the same time as flow rate integration.
● Total rate: Total amount per pulse.
A pulse is outputted when the total volume has attained an amount set by the total rate, and adds to the total pulse count (during total
pulse indication). Settable range: 0.000001 to 99999999
* Set the mass total unit before setting the mass integrating factor. (See Section 4.7.5)
● Regarding pulse width and setup restriction, refer to "4.8.2.1 Total flow pulse (total flow rate, pulse
<Note> Total pulse is not output when TOTAL RATE is set to "0".
<Note> Set the total mode to STOP before start setting.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
Set the mass integrating factor to 100 kg/pulse. * Set the mass total unit beforehand.
(example)
* Setting of the pulse width is the same as described in "4.8.2.1 Total flow pulse (total flow rate, pulse
therefore the description has been omitted.
Key operation
Press the
Press the
ENT
Press the
Press the
ENT
Press the
Press the
ENT
Press the
* Note that, it cannot be entered on the first digit (leftmost).
Press the
Press the
ENT
――― "MASS TOTAL RATE" has been registered. ―――
Press the
Press the
ENT
Press the
Press the
ENT
Press the
ESC
enter the measurement mode.
Description
key twice to display "OUTPUT SETUP".
key once to display "ZERO ADJUSTMENT".
ENT
key for 6 times to display "TOTAL".
key once to display "TOTAL MODE".
ENT
key for 3 times to display "MASS TOTAL RATE".
key once to display the cursor.
ENT
key for 6 times to move the cursor.
key once to display "1".
key once to register.
ENT
key for 5 times to display "TOTAL MODE".
key once to blink the cursor.
ENT
key once to select "TOTAL RESET".
key once to register.
ENT
――― "TOTAL MODE" has been registered. ―――
key twice and then press the
ESC
key for 3 times to
-49-
width)
".
width)
Display
OUTPUT SETUP
ZERO ADJUSTMENT
TOTAL
TOTAL MODE
STOP
MASS TOTAL RATE
0
MASS TOTAL RATE
00
0000000
MASS TOTAL RATE
0
0
00000
00
MASS TOTAL RATE
0
1
00000
00
MASS TOTAL RATE
 COMPLETE 
MASS TOTAL RATE
0
00000100
TOTAL MODE
TOTAL MODE
TOTAL MODE
TOTAL RESET
TOTAL MODE
 COMPLETE 
TOTAL MODE
TOTAL RESET
0.000
m/s
0.000
m3/h
", and
CLEAR
kg
kg
kg
kg
kg
STOP
STOP