4.8.1.3. The source of analog output
Description
● Specify whether volume flow rate (incl. flow velocity) or mass flow rate is output for analog output AO.
AO output: Volume flow rate (factory default), mass flow rate
<Note> The LCD display shows both volume flow rate and mass flow rate.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
Set mass flow rate for analog output AO.
(example)
Key operation
Press the
▼
Press the
ENT
▼
Press the
▼
Press the
ENT
▼
Press the
▼
Press the
ENT
▼
――― "AO OUT.SOURCE" has been registered. ―――
Press the
ESC
the measurement mode.
Description
key twice to display "OUTPUT SETUP".
key once to display "ZERO ADJUSTMENT".
ENT
key for 5 times to display "AO OUT.SOURCE".
key once to blink the cursor.
ENT
key once to display "MASS FLOW RATE".
key once to register.
ENT
key once and then press the
ESC
key for 3 times to enter
-42-
Display
OUTPUT SETUP
ZERO ADJUSTMENT
AO OUT.SOURCE
VOLUME FLOW RATE
AO OUT.SOURCE
VOLUME FLOW RATE
AO OUT.SOURCE
MASS FLOW RATE
AO OUT.SOURCE
COMPLETE
↓
AO OUT.SOURCE
MASS FLOW RATE
0.000
m/s
0.000
m3/h
CLEAR