Displaying The Data In Maintenance Mode - Fuji Electric FSJ Instruction Manual

Ultrasonic flowmeter for steam
Table of Contents

Advertisement

Displaying the data in maintenance mode

Follow the procedure shown below to check possible display errors.
Key operation
Press the
Press the
ENT
Press the
Press the
ENT
 Displays the transit time and the window value calculated from the
piping setting.
Press the
 Displays the measurement value of transit time, T1 (forward time), and
T2 (reverse time) from the piping setting.
Press the
 Displays the measurement value of average transit time, T0, and
transit time difference, DT.
Press the
 Displays the calculated value of pass time of the substances other
than fluid, Ta, and angle of incidence of the fluid,  .
Press the
 Displays the calculation value of sound velocity in fluid, Cf, and
Reynolds number, Re.
Press the
 Displays the analog input pressure AIp and analog input temperature
AIt.
Displays the calculation value selected with the AI range. (See section
4.10.1)
Press the
 Displays the temperature input Pt and density ρ calculation values.
Press the
 Displays the flow velocity distribution correction coefficient K and flow
velocity V1.
Press the
 Displays flow velocity V2 and flow velocity V3.
Press the
 Displays the intensity of received signals.
The larger the value, the larger the intensity of received signals.
Normal measurement values fall in 43% or more.
If the display appears as 0%, no signals are being received.
It is likely that ultrasonic propagation is not possible due to reasons
such as pipe rust.
Press the
Displays receipt signal and noise SNR.
Press the
Displays the noise peak value.
Description
key for 4 times to display "MAINTENANCE MODE".
key once to display "RAS".
ENT
key for 17 times to display "DATA DISPLAY".
key.
ENT
key once.
key once.
key once.
key once.
key once.
key once.
key once.
key once.
key once.
key once.
key once.
-115-
表示
MAINTENANCE MODE
RAS 0000H
0000000000000000
DATA DISPLAY
T0 C:
89 usec
WinC:
80 usec
T1:
0.000 usec
T2:
0.000 usec
T0:
0.000 usec
DT:
0.00 nsec
Ta:
0.0000 usec
 f:
0.000°
Cf:
0.0 m/s
Re:
0
Alp:
0.0 MPa(G)
°
Alt:
0.0
C
Pt:
0.0 °C
ρ: 0.000 kg/m3
K:
1.3333
V1:
0.000 m/s
V2:
0.000 m/s
V3:
0.000 m/s
AGC U:
0.00 %
AGC D:
0.00 %
SNR U:
0. 000 dB
SNR D:
0. 000 dB
Nop U:
000
Nop D:
000

Advertisement

Table of Contents
loading

This manual is also suitable for:

FsxFly