CAUTION ON INSTALLATION LOCATION

(1) A place where ambient temperature and humidity are -20 to +60°C and 95% RH or less for flow
transmitter (FSJ).
(2) Indoor or outdoor locations where there is no exposure to direct sunlight or wind and rain.
(3) A place that provides enough space for periodic inspection and wiring work.
(4) A place not subjected to radiated heat from a heating furnace, etc.
(5) A place not subjected to corrosive atmosphere.
(6) A place not to be submerged.
(7) A place free from excessive vibration, dust, dirt, and moisture.
(8) A place remote from electrical devices (motor, transformer, etc.) which generate electromagnetic
induction noise, electrostatic noise, etc.
(9) A place not subjected to excessive fluid pulsation such as pump discharge side.
(10) A place that provides enough place for the length of the straight pipe.
(11) Altitude: up to 2000 m
(12)
Installation category: II
(13) Pollution degree: 2
Stipulated by IEC61010-1.