Power Meter Operation

Configuring the Remote Interface

Configuring the Remote Interface
This section explains how to configure the power meter for remote control using
either the GP-IB parallel interface or the RS232/422 serial interface. For further
information on using the power meter remotely, refer to the Agilent E4418B/E4419B
Programming Guide.

GP-IB

To select GP-IB as the remote interface press
Select Interface GPIB
GP-IB Address
Each device on the GP-IB (IEEE-488) interface must have a unique address. You can
set the power meter's address to any value between 0 and 30. The address is set to 13
when the power meter is shipped from the factory.
The address is stored in non-volatile memory, and does not change when the power is
switched off, or after a remote interface reset.
Your GP-IB bus controller has its own address. Avoid using the bus controller's
address for any instrument on the interface bus.
To set the GP-IB address from the front panel:
System
1.
Press
Inputs
. The current setting of the GP-IB address is displayed under the
GPIB
GP-IB Addr
2.
To change this setting press
address in a pop up window. Modify this address (see below) as desired.
Use
n
positioned.
Use
n
3.
To confirm your choice press
To set the GP-IB address from the remote interface use the:
SYSTem:COMMunicate:GPIB:ADDRess command.
2-52
System
Inputs
,
(IEEE 488).
,
Remote Interface Configure Interface
softkey.
GP-IB Addr
or
to modify the digit on which the cursor is currently
or
to move to other digits.
.
Enter
,
Remote Interface
,
. The power meter displays the
Agilent E4418B User's Guide
,
,