12.3 Setting the Measurement Conditions
12-10
• Setting the Predefined Rated Voltage
(valid when the assignment method is Set)
12.
Press the cursor keys to select Un Set.
13.
Press SET to display the entry box.
14.
Press the cursor keys to set the predefined value.
15.
Press SET or ESC to close the box.
When the measurement
method is Flicker
• Setting the Measurement Target Frequency
16.
Press the cursor keys to select Frequency.
17.
Press SET to select 50Hz or 60Hz.
When the measurement
method is Flicker
If the measurement method is set to Flicker (normal flicker measurement), set the
measurement target voltage, single observation period, measurement count, and steady-
state range according to the procedure below.
• Setting the Measurement Target Voltage (IEC 61000-4-15 Ed2.0 Only)
18.
Press the cursor keys to select Voltage.
19.
Press SET to select 230V or 120V.
• Setting the Single Observation Period
20.
Press the cursor keys to select the min or sec box by Interval.
21.
Press SET to display the entry box.
22.
Press the cursor keys to set the minutes or seconds.
23.
Press SET or ESC to close the box.
24.
Repeat steps 20 and 23 to enter the minutes and seconds.
When the measurement
method is dmax
When the measurement
method is dmax