12.6 Executing the Measurement of dmax Caused by Manual Switching
• Releasing the Hold Operation
14.
Press HOLD while the display is held. The HOLD key turns OFF, and the display
is updated. In the hold condition, the display can be updated to the most recent
value by carrying out single measurements (by pressing SINGLE).
• Re-measuring the Measurement of dmax Caused by Manual Switching
If a measurement of a given observation period is not performed correctly, you can
change the observation period to be measured by carrying out to the procedure below
and redo the measurement.
15.
Press the Move Period soft key to display the Move Period menu.
16.
Press the cursor keys to set No. (Move to) of the observation period you want to
re-measure.
17.
Press the Move Period Exec soft key. The asterisk moves by the observation
period number specified in step 16.
18.
Press ESC to close the Move Period menu.
19.
Carry out steps 9 to 11 to measure dmax.
If you start the re-measurement, the measurement count and graph display
shown in Count decrease by one. In addition, the total elapsed measurement time
that is displayed to the right of the flicker measurement status decreases by one
observation period. When the re-measurement is complete, the asterisk moves
to the next observation period number of the observation period in which the
measurement is completed.
20.
If you want to continue with the measurement, repeat steps 9 to 11. To change the
observation period to be measured, return to step 15.
• Aborting the Measurement of dmax Caused by Manual Switching
21.
Press the Reset soft key. An Alert dialog box appears.
22.
Press the cursor keys to select OK.
23.
Press SET. The measurement of dmax caused by manual switching is aborted.
The measured data and judgement up to that point are discarded. The flicker
measurement status indicates Reset. To continue the measurement of dmax
caused by manual switching, select Cancel in step 22 and press SET.
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
12-27