Voltage Fluctuation and Flicker Measurement (Option)
12.1 Voltage Fluctuation and Flicker Measurement
Function
IM WT3001E-51EN
This function allows you to perform voltage fluctuation and flicker measurements
conforming to IEC 61000-3-3. If you use the Harmonic/Flicker Measurement Software
sold separately, you can print a report of the measurement judgement results.
• Connecting to the Harmonic/Flicker Measurement Software
For details on connecting this instrument to the Harmonic/Flicker Measurement
Software, see the user's manual of the Harmonic/Flicker Measurement Software.
When you use the Harmonic/Flicker Measurement Software, this instrument enters
remote mode in which key operations on this instrument are disabled. The settings
described in the subsequent sections are set automatically or set from the setup
screen of the software.
Note
Do not release the remote mode of this instrument and change the settings while operating this
instrument with the Harmonic/Flicker Measurement Software. Doing so may impede voltage
fluctuation and flicker measurements conforming to the IEC standard. If you release the remote
mode of this instrument, close the Harmonic/Flicker Measurement Software.
• Measurement Items (Measurement Functions)
Measurement Items
Rated voltage Un
Voltage frequency Freq
Relative steady-state voltage change dc
Maximum relative voltage change dmax
Period during which relative voltage change exceeds
*3
the threshold level Tmax
Short-term flicker value Pst
Long-term flicker value Plt
Instantaneous flicker sensation (IFS)
Cumulative probability function (CPF)
*1 The limit is defined only for 230 V/50 Hz.
*2 The limits varied depending on the EUT.
*3 Tmax for IEC 61000-3-3 Edition 3.0, d(t) for IEC 61000-3-3 Edition 2.0.
This chapter describes about the example when setting the edition number of IEC61000-3-
3 as Ed3.0. When setting the edition number of IEC61000-3-3 as Ed2.0, please paraphrase
Tmax in this chapter as d(t).
*4 The limits of threshold level is 3.3%.
*5 Cannot be displayed on this instrument screen. The value can be displayed on the screen of
the Harmonic/Flicker Measurement Software.
• Measurement Method
The following two voltage fluctuation and flicker measurement methods are available.
• Normal flicker measurement
Calculates all the values of dc, dmax, Tmax, Pst, and Plt, compares them to the
preset limits, and indicates the total judgement.
• Measurement of dmax caused by manual switching
Measures the maximum relative voltage change, dmax, when the EUT switch is
manually turned ON and OFF, determines the average over 24 measurements, and
compares and judges against the limit.
Limit
3.3% or less
4% or less
6% or less
7% or less
500 ms or less
1.0 or less
0.65 or less
*5
*5
1
2
3
4
5
6
7
*1
8
*2
*2
*2
*4
9
10
11
12
13