Explanation
IM WT3001E-51EN
• Setting the Measurement Count
25.
Press the cursor keys to select Count.
26.
Press SET to display the entry box.
27.
Press the cursor keys to set the measurement count.
28.
Press SET or ESC to close the box.
• Setting the Steady-State Range (dmin: Allowable Range of Relative
Voltage Change to Be Considered Steady-State)
29.
Press the cursor keys to select dmin.
30.
Press SET. An entry box appears.
31.
Press the cursor keys to set the steady-state range.
32.
Press SET or ESC to close the box.
For an explanation of the terminology, see section 12.1.
• Measurement Method
Select the voltage fluctuation and flicker measurement method from the two methods below.
• Flicker (normal flicker measurement)
Judges the voltage fluctuation and flicker from dc, dmax, Tmax, and Pst.
• dmax (measurement of dmax caused by manual switching)
You manually turn the EUT switch ON. This instrument measures the voltage
fluctuation caused by the inrush current that flows when the power is turned ON,
and judges the voltage fluctuation and flicker from the dmax average.
• Edition Number of IEC 61000-4-15
You can select the edition number of IEC 61000-4-15.
• Edition 1.1
• Edition 2.0
IEC 61000-4-15 specifies requirements for measurement instruments.