IM WT3001E-51EN
• Setting the Edition Number of IEC 61000-4-15
3.
Press the Flicker Settings soft key to display the Flicker Settings dialog box.
4.
Press the cursor keys to select 4-15 of IEC 61000 Edition.
5.
Press SET to select Ed1.1 or Ed2.0 of IEC 61000-4-15.
When the measurement
method is Flicker
• Setting the Edition Number of IEC 61000-3-3
6.
Press the cursor keys to select 3-3 of IEC 61000 Edition.
7.
Press SET to select Ed2.0 or Ed3.0 of IEC 61000-3-3.
When the measurement
method is Flicker
• Setting the Measured Element
8.
Press the cursor keys to select the element you want to set from Object.
9.
Press SET. The button to the left of the selected element is highlighted indicating
that voltage fluctuation and flicker measurement will be performed on the element.
To remove the element from the measurement, press SET again.
The highlighting of the button is cleared.
When the measurement
method is Flicker
• Setting the Rated Voltage
• Selecting the Assignment Method of the Rated Voltage
10.
Press the cursor keys to select Un Mode.
11.
Press SET to select Auto or Set.
If you select Auto, proceed to step 16.
If you select Set, proceed to step 12.
12.3 Setting the Measurement Conditions
When the measurement
method is dmax
When the measurement
method is dmax
When the measurement
method is dmax
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
12-9