5.7 Setting of the Time Difference from GMT or SNTP
Explanation
5-24
• Setting SNTP (Simple Network Time Protocol)
10.
Press the cursor keys to select Exec at Power On under SNTP.
11.
Press SET to select ON or OFF.
12.
Press the cursor keys to select Server.
13
Press SET to display the keyboard.
14.
Use the keyboard to enter the IP address of the SNTP server.
15.
Press ESC to close the dialog box.
16.
Press the cursor keys to select Time Out.
17.
Press SET to display the entry box.
18.
Press the cursor keys to set the timeout value in the range of 1 to 60 (s).
19.
Press SET to confirm.
• Restarting this instrument
If Exec at Power On is ON, restart this instrument to synchronize the time with the
SNTP server.
Set the time difference from GMT (Greenwich Mean Time).
• Setting of the time difference from GMT
Set the time difference in the range of –12 hours 00 minutes to 13 hours 00 minutes.
For example, Japan standard time is ahead of GMT by 9 hours. In this case, set Time
Hour to 9 and Minute to 00.
• Checking the Standard Time
Check the standard time for the region where this instrument is to be used using one
of the following methods.
• Check the date and time settings on your PC (Windows).
• Check the site at the following URL:http://www.worldtimeserver.com/
Note
This instrument does not support Daylight Savings Time. To set the Daylight Savings Time,
adjust the time difference from GMT.
• Setting SNTP
If Exec at Power On is ON, this instrument synchronizes with the SNTP server clock
that is connected to the network when the power is turned ON.
Note
If the time difference from GMT (Greenwich Mean Time) is specified, the time is set to a
value derived by calculating the difference between the time retrieved from the SNTP server
and the time difference.
It is possible to synchronize with the time on an SNTP server when setting the date/time of
this instrument. For details, see section 3.12 in the User's Manual IM WT3001E-01EN.
If you do not want to synchronize with the time on an SNTP server, do not set the IP address