YOKOGAWA WT3002E User Manual page 13

Precision power analyzer
Hide thumbs Also See for WT3002E:
Table of Contents

Advertisement

Explanation
IM WT3001E-51EN
1.3 Selecting the Type of the Revolution and Torque Signals
• Selecting the Torque Signal Type
2.
Press the cursor keys to select the Sense Type under Torque.
3.
Press SET. A signal type selection box appears.
4.
Press the cursor keys to select Analog or Pulse.
5.
Press SET to confirm the signal type.
• Selecting the Signal Type
The signal applied from the revolution sensor or torque meter to this instrument can
be selected from the following two types.
• Analog
Select this when the signal type is a DC voltage (analog signal).
• Pulse
Select this when the signal type is a pulse signal.
• Signal Type and Setup Items
Some of the settings described in the subsequent sections are not required depending
on the signal type as described below.
• Settings Related to the Revolution Signal Type
Analog range (section 1.4)
Line filter (section 1.5)
Synchronization source (section 1.5)
Pulse range (section 1.6)
Pulse count per revolution (section 1.6)
* Measurement is possible using None (default), but it is recommended that you specify the
setting to improve the measurement accuracy. For details, see the explanation in section
1.5.
• Settings Related to the Torque Signal Type
Analog range (section 1.4)
Line filter (section 1.5)
Synchronization source (section 1.5)
Pulse range (section 1.6)
Pulse rating (section 1.6)
* Measurement is possible using None (default), but it is recommended that you specify the
setting to improve the measurement accuracy. For details, see the explanation in section
1.5.
Revolution Signal Type
Analog
Pulse
Required
Not required
Required
Not required
Required
Recommended*
Not required
Required
Not required
Required
Torque Signal Type
Analog
Pulse
Required
Not required
Required
Not required
Required
Recommended*
Not required
Required
Not required
Required