Emergency recovery from power failure

Emergency recovery from power failure
1. Rotation of the rotor
The rotor stops as a result of free coast. However, if the rotor did not decelerate to 300 rpm or lower
when a power failure occurred, it will, or lower when a power failure occurred, after power returns,
automatically accelerate again and return to the set speed before the power failure. If the speed goes
below 300 rpm, the rotor will decelerate and stop after power returns.
If the rotor is stopping when the power returns, this micro-ultracentrifuge does not accept the
VACUUM button etc. (see
2. Touchscreen display
All displays are off during a power failure. After power returns, the instrument starts control again at
the settings specified before the power failure. (The settings have battery-backed back-ups.)
The instrument then displays an alarm message, indicating that it has suffered a power failure.
3. Removing the rotor from the micro-ultracentrifuge
2-70
Sorvall MTX 150
The operation history of the micro-ultracentrifuge can be output
in CSV format to a USB memory on the market.
Figure 2-24. USB/LAN communication setting screen
DANGER Before removing the front cover for troubleshooting, always turn off the
POWER switch of the instrument, unplug the power cord from the wall outlet, and wait
at least three minutes to avoid the risk of electrical shock.
WARNING
(1) Never open the door during rotation.
(2) Never touch the rotor during rotation.
CAUTION
Never conduct operations in a manner other than as described in this operation manual.
Contact with your nearest service representative if any problem arises.
"Alarm indicators"
*CSV format: This is a versatile file format that is compatible with
spreadsheet softwares such as EXCEL.
on
page
4-2).
Press the DATA EXPORT button.
(Writing in the USB memory is
automatically started.)