Calibration And Setting; Zero Calibration - Fuji Electric ZSS-8 Series Instruction Manual

Cross stack laser gas analyzer
Table of Contents

Advertisement

6. CALIBRATION AND SETTING

Select "Zero Calibration" from the "Menu" screen and press the
6.1.1 Preparation
AUTION
Parts name
Calibration gas cell
Cable between receiver unit and
transmitter unit
Cable between receiver unit and
control unit (for calibration)
Zero gas (N
)
2
Pressure regulator
Pipe (PTFE tube, etc)
Joint (Rc1/4)
Flow meter
Thermometer
Pressure gauge
Others (joint, etc)
(1) Turn OFF the power.
(2) Remove the receiver box and the transmitter box using the hexagon wrench. Never remove the boxes
while the system into which the analyzer is incorporated is in operation. Otherwise, hot temperature gas
may blow out.

6.1 Zero calibration

 Do not detach the transmitter unit during calibration while the pow-
er is supplied. If you turn the transmitter unit towards a person, the
laser beam can damage his/her cornea.
 The laser beam is the invisible infrared light. Do not watch the laser
beam directly or scattering light.
 Do not watch the laser beam directly with the optical measuring
device. Otherwise, it may cause serious damage to your eyes.
Quantity
1
1
1
1
1
Several m
2
1
1
1
- 42 -
key.
NT
Remarks
To be ordered separately
To be ordered separately
To be ordered separately
To be ordered separately
To be ordered separately
10/8 mm or larger
For connecting the zero gas line and the
exhaust line to the calibration gas cell
2L/min or more
4 to 20mA output, for temperature correc-
tion (required for range HCl and NH
4 to 20mA output, for pressure correction
(required for low range HCl and NH
)
3
)
3
INZ-TN8ZSS-E