6. SETTING AND CALIBRATION

6.1 Switch of range

6.1.1 Setting of range switch mode

ENT
ed.
Description of setting
MR: Select a desired range on this screen.
RR: Select a desired range according to
the remote range switch contact input.
AR: Automatically switched from Range 1
to Range 2 when the measured concen-
tration exceeds 90% of Range 1.
Automatically switched from Range 2
to Range 1 when the measured concen-
tration becomes less than 80% of
Range 1.
* Operation set for each Ch only can be
performed.
INZ-TN2ZPB-E
MODE
key.
ENT
key.
ENT
Measurement Mode Screen
User Mode
Switch Ranges
Calibration Parameters
Alarm Setting
Setting of Auto Calibration
Setting of Auto Zero Calibration
Peak Alarm Setting
Parameter Setting
-
Switch Ranges
-
Ch1
MR
NOx
Ch2
AR
SO
2
Ch3
RR
CO
2
Ch4
MR
CO
Ch5
MR
O
2
-
-
Switch Ranges
Ch1
MR
NOx
Ch2
AR
SO
2
Ch3
RR
CO
2
Ch4
MR
CO
Ch5
MR
O
2
or previous screen
25
MODE
Select an item
with UP/DOWN and ENT
Back with ESC
ENT
Select Ch No.
with UP/DOWN and ENT
Back with ESC
Range1
0–50.00
Range2
0–500.0
Range1
0–50.00
Range2
0–500.0
Range1
0–10.00
Range2
0–20.00
Range1
0–50.00
Range2
0–500.0
Range1
0–10.00
Range2
0–25.00
ENT
(
)
Select Ch No.
Switch ranges
with UP/DOWN and ENT
Back with ESC
Range1
0–50.00
Range2
0–500.0
Range1
0–50.00
Range2
0–500.0
Range1
0–10.00
Range2
0–20.00
Range1
0–50.00
Range2
0–500.0
Range1
0–10.00
Range2
0–25.00
ENT
Range switch
ppm
ppm
ppm
ppm
vol%
vol%
ppm
ppm
vol%
vol%
ppm
ppm
ppm
ppm
vol%
vol%
ppm
ppm
vol%
vol%

Advertisement

Table of Contents
loading

Table of Contents